Syringe Size
|
10, 20, 30, 50/60 ml
|
Applicable Syringe
|
Compatible with syringe of any standard
|
Flow Rate
|
Syringe 10 ml: 0.1-400 ml/h
Syringe 20 ml:0.1-600ml/h <100 ml/h in 0.1ml/h increments
Syringe 30 ml:0.1-900ml/h ≥100 ml/h in 1ml/h increments
Syringe 50/60 ml: 0.1-1300ml/h
|
Bolus Rate
|
Syringe 10 ml: 400 ml/h
Syringe 20 ml: 600 ml/h
Syringe 30 ml: 900 ml/h Syringe 50/60 ml: 1300 ml/h
|
Anti-Bolus
|
Automatic
|
Accuracy
|
±2% (mechanical accuracy ≤1%)
|
Infusion Mode
|
Flow rate: ml/min, ml/h Time-based
Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h etc.
|
KVO Rate
|
0.1-1 ml/h (in 0.1 ml/h increments)
|
Alarms
|
Occlusion, near empty, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, reminder alarm, pressure sensor error, syringe installation error, syringe drop off
|
Additional Features
|
Real-time infused volume, automatic power switching,
automatic syringe identification, mute key, purge, bolus, anti-bolus, system memory
|
Drug Library
|
Available
|
Occlusion Sensitivity
|
High, medium, low
|
Power Supply, AC
|
110/230 V (optional), 50/60 Hz, 20 VA
|
Battery
|
9.6±1.6 V, rechargeable
|
Battery Life
|
7 hours at 5 ml/h
|
Working Temperature
|
5-40?
|
Relative Humidity
|
20-90%
|
Atmospheric Pressure
|
860-1060 hpa
|
Safety Classification
|
Class ?, type CF
|